Decay and activity, planned.

Pick a radionuclide or type a half-life, enter an activity and a time, and read the activity at the target time, or what a calibration must read to have that activity later.

Last updated September 17, 2026

This planner applies the exponential decay law to a single radionuclide: the activity left after any time from a calibration, or the reverse, for the 27 clinical radionuclides of the isotope table or any custom half-life. It is for people who plan radiopharmaceutical work, radiochemists, imaging scientists and students, and it prints the formula, the half-life it used and the same activity in MBq, GBq, mCi and µCi with every result.

Planner

27 nuclides of the isotope table (ENSDF half-lives), or a custom half-life.

Direction

The activity on the calibration certificate, or measured now.

Elapsed time; a whole number of half-lives halves the activity that many times.

Results

Activity at target
5.411GBq
MBq
5411
GBq
5.411
mCi
146.3
µCi
146300
Elapsed half-lives
0.4515
Fraction remaining
73.13%
Decay constant λ
0.004347h⁻¹
Half-life used
6.6443 d · ¹⁷⁷Lu, ENSDF · 574,068 s

For reference and education. Nothing on this page is clinical guidance, a dose recommendation or an Epitorix performance figure; check every number against your own validated procedures.

The formulas

ln 2 = 0.693 147

One exponential

A(t)
A(t) = A₀ · e^(−λt)
The activity after time t from an initial activity A₀. Activity is the number of decays per second; it falls at the same rate as the number of atoms.
λ
λ = ln 2 / T½
The decay constant, the fraction of atoms that decay per unit time. ¹⁷⁷Lu: 0.004 347 h⁻¹; ¹⁸F: 0.3789 h⁻¹.
n, fraction
n = t / T½ ; fraction = 2^(−n)
Elapsed half-lives and the fraction left: 1 half-life leaves 50 %, 2 leave 25 %, 3.32 leave 10 %, 10 leave 0.098 %.
Reverse
A₀ = A(t) · e^(λt) = A(t) · 2^(n)
What the calibration must read to have A(t) at the target time. The same law, run backwards.
Curie
1 Ci = 3.7 × 10¹⁰ Bq (exact) ; 1 mCi = 37 MBq ; 1 GBq = 27.03 mCi
The becquerel is the SI unit (one decay per second); the curie is defined in terms of it. The conversion is exact and independent of the nuclide.

Worked examples

ENSDF half-lives

By hand

Each row is computed from the formulas above with the isotope table's half-lives; the planner reproduces every value.

Worked examples of the decay law with the isotope table's half-lives.
Nuclide · T½InputTimeHalf-livesResultNote
¹⁷⁷Lu · 6.6443 d7.4 GBq at calibration6.647 d1.00043.699 GBqone ICRP-107 half-life later: 200 mCi has become 100 mCi
¹⁸F · 109.77 min370 MBq at calibration109.77 min1.0000185 MBqexactly one half-life: 10 mCi → 5 mCi
⁹⁹ᵐTc · 6.0072 h740 MBq at calibration6.0072 h1.0000370 MBq20 mCi → 10 mCi
⁶⁸Ga · 67.71 min1 000 MBq at calibration30 min0.4431735.6 MBq2^(−0.4431) = 0.7356
¹⁷⁷Lu · 6.6443 d7.4 GBq wanted at target3 d0.451510.12 GBqreverse: the calibration must read 2^0.4515 = 1.3675 times more
²²⁵Ac · 9.920 d10 MBq wanted at target3 d0.302412.33 MBqreverse; parent activity only, the chain is not modeled
custom · 10 h100 MBq at calibration30 h3.000012.5 MBqthree half-lives: 1/8

Questions

What the numbers mean

What is a half-life?

The time in which the activity of a radionuclide, and the number of its atoms, falls by half. After n half-lives the fraction left is 2 to the power of minus n: 50 %, 25 %, 12.5 % and so on. It is a property of the nuclide alone, independent of the amount, the temperature and the chemistry it is in.

Where do the half-lives come from?

From the isotope table on this site: ENSDF evaluations read through NuDat 3 and the IAEA Live Chart API on 2026-09-16, with the evaluation cut-off recorded per nuclide. ICRP-107 (2008) differs in the third or fourth figure for some nuclides, ¹⁷⁷Lu 6.647 d against 6.6443 d for instance; over one half-life that difference moves the activity by less than 0.05 %.

Why both MBq and mCi?

The becquerel is the SI unit of activity, one decay per second, and the unit European labels and most journals use. The curie is defined as exactly 3.7 × 10¹⁰ Bq and remains the working unit in US practice, so 1 mCi = 37 MBq and 1 GBq = 27.03 mCi. The conversion is exact; nothing about the nuclide enters it.

Does the planner account for daughters, residuals or dose?

No. It applies the single-exponential law to the parent's activity only. Decay chains with their own half-lives (²²⁵Ac, ²¹²Pb, ²²³Ra), the activity left in a syringe or tubing, and any question of absorbed dose or dosimetry are outside it and belong to your validated procedures.

Sources

A reference tool. It evaluates one exponential from a published half-life; it does not plan, prescribe or verify a dose, and nothing here is an Epitorix performance figure.